EDAN iM80 User Manual - page 80
Patient Monitor User Manual Monitoring RESP
- 69 -
from the pop-up list.
9.9 Changing Hold Type
To change the calculation mode, in the Resp Setup menu, set Hold Type to Manual or Auto.
When it is set to the AUTO mode, Hold High and Hold Low are unavailable, and the monitor
can calculate the respiration rate automatically. When it is set to the Manual mode, you can
adjust the broken lines in RESP area by the Hold High and Hold Low items.
9.10 Changing the Size of the Respiration Wave
Select the Resp waveform area to open the Resp Wave Setup menu:
Select AMP, then choose an apprioate value. The value is bigger, the waveform amplitude is
higher.
Select Sweep: select an apprioate setting from the pop-up list.
9.11 Using Resp Alarms
Resp alarms can be switched on and off and the high and low alarm limits can be changed just
like other measurement alarms, as described in the Alarms chapter.
9.12 Changing the Apnea Time
The apnea alarm is a high priority red alarm used to detect apneas. The apnea alarm delay time
defines the time period between the point where the monitor cannot detect any respiration activity
and the indication of the apnea alarm.
1. In the Resp Setup menu, select Apnea Alm.
2. Select the appropriate setting from the popup list.