Patient Monitor User Manual EMC Information
- 168 -
IEC/EN 61000-4-4
±1 kV for input /output
signal
±1 kV for input
/output signal
typical commercial or
hospital environment.
Surge
IEC/EN 61000-4-5
±1 kV for line to line
±2 kV for line to ground
±1 kV for line to
line
±2 kV for line to
ground
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power frequency
(50/60Hz)
magnetic field
IEC/EN 61000-4-8
3A/m
3A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC/EN
61000-4-11
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
T
(>95% dip in U
T
)
for 5 sec
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
T
(>95% dip in U
T
)
for 5 sec
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the Patient
Monitor requires
continued operation
during power mains
interruptions, it is
recommended that the
Patient Monitor be
powered from an
uninterruptible power
supply or a battery.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
B.3 Electromagnetic Immunity - for EQUIPMENT and SYSTEMS
that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
The Patient Monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of Patient Monitor should assure that it is used in such an