EDAN iM80 User Manual - page 167
Patient Monitor User Manual Product Specification
- 156 -
Alarm Type
SYS, DIA, MAP
Pressure Resolution
1mmHg
Measuring Accuracy
Maximum Mean Error
±5mmHg
Maximum
Standard
Deviation
8mmHg
Adult/Pediatric
Neonate
Normal
Condition
Single Fault
Condition
Normal
Condition
Single fault
Condition
Cuff Pressure
Measuring Range
300 mmHg
330 mmHg
150 mmHg
165 mmHg
Maximum
Measuring
Period
Less than 160s
Less than 180s Less than 80s Less than 90s
A.11 SpO
2
A.11.1 SpO
2
from EDAN Module
Measuring Range
0 ~ 100 %
Resolution
1 %
Accuracy
Adult (including Pediatric)
±2 % (70%~100% SpO
2
)
Undefined (0~69% SpO
2
)
Neonate
±3 % (70%~100% SpO
2
)
Undefined (0~69% SpO
2
)
Pulse Rate
Measuring Range
25bpm ~ 300bpm
Resolution
1bpm
Accuracy
±2bpm
Data update period
1s
Wave length
Red light
660
±3 nm
Infrared light
905
±5 nm
Emitted light energy
≤15 mW