EDAN iM80 User Manual - page 50
Patient Monitor User Manual Alarm Information
- 39 -
CO
2
Check Adapter
The cannula is off or
disconnected.
Low
AA Out Range
The AA concentration
exceeds the accuracy range
of AG module.
High
Reduce AA
concentration.
O
2
Sensor Error
The oxygen sensor of the
sidestream AG module has
a failure.
Medium
Stop measuring of
AG
module, and
notify biomedical
engineer or
Manufacturer’s
service staff.
QuickTemp Comm
Fail
TEMP module failure or
communication failure.
High
Stop measuring
function of TEMP
module, and notify
biomedical engineer
or
manufacturer’s
service staff.
Temp exceed limit
The TEMP value is beyond
the range of +25
°C ~
+45
°C.
Med
Put the sensor into
the sensor bracket,
take it out and
measure again.
No Temp Sensor
TEMP sensor is not
connected to the TEMP
module.
Low
Connect the sensor
and the monitor well,
and measure again.
Ambient temp too
high
The Sensor temperature is
higher than +40
°C.
Low
Put the sensor into
the sensor bracket,
measure again after
the ambient
temperature reaches
normal value.
Ambient temp too low
The Sensor temperature is
lower than +10
°C.
Low
Put the sensor into
the sensor bracket,
measure again after
the ambient
temperature reaches
normal value.