If the replacement head was not factory‐calibrated, the following
warning screen appears.
Figure 7-58: Invalid-data warning screen
If this happens:
a. Remove the head and replace it with another one.
b. In the wizard screen, select the check box to confirm that you have
installed another print head, and click
Next
.
c. Contact your authorized Objet Customer Support center about the
unformatted head.
•
If the vacuum test is not successful, the replacement head was not
sealed properly during installation, and the following warning screen
appears.
Figure 7-59: Vacuum Leakage warning screen
If this happens:
a. Click
Next
.
b. Follow the instructions on the wizard screens to re‐install the head.