Operating & Maintaining the Objet30 3-D Printer
7–44
DOC-24000 Rev. C
Weight Calibration
After the successful installation of print heads, the wizard begins the
weight test, in which each head prints a sample. This test takes 15–25
minutes.
Figure 7-60: Weight test progress screen
14. Carefully remove the samples for each head, keeping track of the head
that printed them.
Figure 7-61: Samples printed on the Objet30 printer build tray
15. Using the scale, weigh each sample and enter the weight (in grams) for
the corresponding head in the wizard screen.
16. Close the printer cover, then click
Next
.
Figure 7-62: Weight Test data entry screen
Support Print Head
Model Print Head